PRODUCT DESCRIPTION 2
2.1.2 Response time
The response time is the time required to attain 90% of the signal peak after a sudden
change in the measurement signal. It can be set anywhere between 1 and 600 s. As the
response time increases, transient measured value fluctuations and interruptions are
damped stronger and stronger which “smoothes out” the output signal.
Fig. 2: Response time
100
98
96
94
92
90
88
86
84
Response time
t in s
Measured value with response time
90% of the signal peak
Process change
0 10 20 30 40 50 60 70 80 90
Measured value
in %