37
8026957/AE00/V1-0/2021-08| SICK OPERATING INSTRUCTIONS | DUSTHUNTER SB30
Subject to change without notice
START-UP AND PARAMETER SETTINGS 4
4.2.2 Changing settings using the jumper on the processor board
The following steps are then necessary (see “Adapting the sender/receiver unit to the duct
diameter”, page 32):
Loosen CAMLOCK clips of the sender/receiver unit, swivel the electronics unit to the
side and take off from the mounting flange.
Loosen the fastening screws on the electronics unit, carefully pull the electronics unit
out of the cover.
After changing the settings, reassemble the sender/receiver unit in reverse sequence.
4.2.2.1 Setting measuring range and limit value
Plug the jumper in the slot for the setting desired for the measuring range and limit
value.
Fig. 22: Processor board setting options
Parameter
Slot 1 Slot 2 Slot 3 (default)
Value at LZ
0 SI 0 SI 0 SI
Value at MBE
100 SI 500 SI 3000 SI
Limit value
50 SI 250 SI 1500 SI
Slot 4 is not used.
Microswitch for baud
rate Modbus
®
Rotary switch for
setting measuring
range
Microswitch for
addressing