MAINTENANCE 6
6.2 Maintenance on the sender/receiver unit
Clean the outside of the sender/receiver unit in regular intervals. Remove deposits with
water or mechanically using suitable auxiliary means.
Work to be performed
▸ Loosen CAMLOCK clips (1) of the sender/receiver unit and swivel electronics unit (2) to
the side.
▸ Check mounting flange (5) and purge air connections (see “Sender/receiver unit DHSB-
T”, page 15) for contamination and clean when necessary.
▸ Carefully clean sender optics (3) and receiver optics (4) with an optics cloth or cotton
swabs.
▸ Reassemble the sender/receiver unit again.
▸ Resume Measuring mode.
Fig. 30: Cleaning the optical interfaces
NOTE:
▸ Do not damage any device parts during maintenance work.
▸ Do not interrupt the purge air supply.
WARNING: Danger from exhaust gas
The sender/receiver unit must be opened for cleaning. It is possible that gas escapes in
plants with overpressure in the duct.
▸ Take suitable protection measures or open the sender/receiver unit only
when the plant is at a standstill.
CAMLOCK clips
Receiver optics
Electronics unit
Mounting flange
Sender optics