8026957/AE00/V1-0/2021-08| SICK OPERATING INSTRUCTIONS | DUSTHUNTER SB30
Subject to change without notice
START-UP AND PARAMETER SETTINGS 4
Loosen fastening screws (4) on the electronics unit, carefully pull the electronics unit out
of cover (5) and loosen plug connector (6) for the measurement receiver from processor
board (7).
Fig. 19: Electronics unit opened
Loosen fastening screws (8) for tube (9) and take the tube off (see “Electronics unit
opened”, page 33).
Unscrew mounting ring (10) from the tube, turn it round and screw it back on again.
Fig. 20: Tube with mounting ring
Set the immersion depth on microswitch 4 (11).

Plug connector
Processor board
Fastening screws
Tube

Tube
Mounting ring