8 Maintenance
8.1 Cleaning
To maintain the reading performance of the TriSpector1000, it is necessary to keep
the camera window and the laser window clean. Check the windows for contamination
such as dust or moisture regularly. Clean with a damp and soft cloth and use a mild
cleaning agent if necessary.
For the TriSpector1000 PMMA variants, the windows can attract dust particles due to
static charge. Use the SICK anti-static plastic cleaning agent (no. 5600006) in combina‐
tion with the SICK lens cloth (no. 4003353) to reduce this effect.
NOTE
No maintenance is required to ensure compliance with the laser class.
8.2 Upgrading the firmware
The latest firmware for the TriSpector1000 can be downloaded from the SICK Support
Portal.
To upgrade the firmware:
1. Log in to the SICK Support Portal.
2. Under the Vision tab, select TriSpector1000 and navigate to the Firmware download for
TriSpector1000 page.
3. Download the firmware package file (FWP file format) and the firmware key file
(KEY file format).
4. Follow the instructions on the page to install the firmware.
NOTICE
Old firmware for the TriSpector1000 is available on the SICK Support Portal and can be
used to downgrade the device. Before downgrading, make sure all application parame‐
ters are reset to default:
1. Click Load application defaults on the Functions menu.
2. Click Save permanent on the Parameter menu.
Failure to reset the parameters might result in a device that does not boot and needs
re-flashing.
8.3 Data backup and restoration
Creating a back-up of a configuration
Use the Export function in SOPAS ET to create a back-up of a TriSpector1000 configura‐
tion and save to a PC, see "Exporting and importing data", page 56. Use the Import func‐
tion to re-import the configuration into SOPAS ET.
Resetting parameters to default
To reset all SOPAS ET parameters including IP address:
1 Click Load factory defaults on the Functions menu.
2 Click Save permanently on the Parameter menu.
To reset all SOPAS ET parameters except IP address:
1 Click Load application defaults on the Functions menu.
2 Click Save permanently on the Parameter menu.
8 MAINTENANCE
58
O P E R A T I N G I N S T R U C T I O N S | TriSpector1000 8021808/12ID/2019-01 | SICK
Subject to change without notice