Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779144002-01, 09/2018
29
Electrical characteristics
Fuse values
1-channel unit (7MB2121, 7MB2127, 7MB2124, 7MB2126)
100 to 120 V
200 to 240 V
2-channel unit (7MB2123, 7MB2128, 7MB2124, 7MB2126)
100 to 120 V
200 to 240 V
1T/250
0.63T/250
1.6 T/250
1T/250
Gas inlet conditions
Permissible sample gas pressure 600 to 1500 hPa absolute,
700 to 1300 hPa absolute with installed pressure switch
Sample gas flow
18 to 90 l/h (0.3 to 1.5 l/min)
Sample gas temperature
0 to 50 °C
Sample gas humidity < 90% RH (relative humidity), non-condensing or
dependent on measuring task
Parts in gas path wetted by sample gas
Standard devices
Gas path with tubes
Gas feedthrough
Stainless steel 1.4571
Hose
FKM
Hose coupling
PA 6
Sample chamber
Body
Lining
Gaskets (O-rings)
Window
Alu
Alu
FKM or FFKM
CaF
2
, adhesive E353
Pipe gas path
Gas feedthrough
Ti or stainless steel 1.4571
Pipe
Ti or stainless steel 1.4571
Sample chamber
Body
Lining
Gaskets (O-rings)
Window
Alu
Alu or Ta
FKM or FFKM
CaF
2
, adhesive E353
Special applications
Pipe gas path
Gas feedthrough
Ti or stainless steel 1.4571 or Hastelloy C22
Pipe
Ti or stainless steel 1.4571 or Hastelloy C22
Sample chamber
Body
Lining
Gaskets (O-rings)
Window
Ti or stainless steel 1.4571 or Hastelloy C22
Ta or Alu
Material according to specification
CaF
2
; without adhesive
Further designs upon request
Options
Flowmeter
Measurement pipe
Duran glass
Variable area
Duran glass
Suspension boundary
PTFE