Installing the CAC3 on the SIB3
It will be neccessary to download the database saved earlier using GCP.
1. Carefully align the CAC3 connectors over the mating connectors of the SIB3.
2. Press down to fully engage the connectors.
3. Install the 2mm screws with o-rings. Avoid overtightening.
4. Plug in the ethernet cable.
5. Follow the steps in the
General Analyzer Startup Procedure
in the
Maxum Edition II
Analyzer General Maintenance Manual.
Removing a SIB3
1. As each cable is disconnected, move it clear of the SIB so it can be removed.
2. Make note of
I/O board positions
I/O cable location
All cable positions to the SIB3
SIB3 I/O Mode switch
3. Disconnect
I/O cables from SIB3 and I/O boards
Display cables
Ethernet cable from CAC3
Internal ethernet cable to SNEs if used
4. Remove ESB
5. Remove I/O boards
6. Disconnect from the SIB3
I
2
C cables
Purge cable
Power cable
Serial cables
CAN bus cable if present
7. If the existing CAC3 will be reused, remove it following the procedure in the preceding
section and place in a static-protected area.
Electronic Compartment Component Descriptions and Maintenance Procedures
2.3 System Controller Version 2.1 (SYSCON2.1)
Maxum II Reference Manual
Manual, 7/2017, 2000596-001 51

Table of Contents

Other manuals for Siemens Maxum II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Maxum II and is the answer not in the manual?

Siemens Maxum II Specifications

General IconGeneral
CategoryGas Chromatograph
Measurement PrincipleGas Chromatography
Communication ProtocolsModbus, Ethernet, Profibus
Operating Temperature-20°C to 55°C
Power Supply24 VDC or 100-240 VAC
Mountingrack mount
ApplicationsRefinery, chemical, petrochemical, natural gas
DetectorsThermal Conductivity Detector (TCD), Flame Ionization Detector (FID)
Sample InletsMultiple
DimensionsVaries by configuration (e.g., 19" rack mount)

Related product manuals