Mechanical systems 10 - 27
Siemens-Elema AB SPR8-215.898.01 Page 27 of 28 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
Install the new friction linings according to (Fig. 31 or Fig. 32). There are two different
vertical friction linings. Both are equally suitable for the unit.
Reinstall the upper counter bearing.
Tighten the screws.
Check that there is no braking action on the counter bearings.
Tighten the adjustment screw. Make sure that the spring mechanism is under tension
before releasing the ceiling support.
Attach the rear, left right and front covers.
Check parallelism, refer to section "Parallelism".
Adjust counterbalance, refer to section "Counterbalance".
Adjust friction linings, refer to section "Adjusting the friction linings".