Capacitor bank 7 - 3
Siemens-Elema AB SPR8-215.898.01 Page 3 of 6 MOBILETT Plus / - E/ - M
Solna, Sweden Rev. 05 01.02 SPS-UD
• Measure the voltage between the + and - bars; the voltage must be less than 10 V.
• Note the cable connections to the capacitor bank (Fig. 3 for MOBILETT Plus/ Plus M and
Fig. 4 for MOBILETT Plus E)
Fig. 3 MOBILETT Plus/ Plus M
Fig. 4 MOBILETT Plus E
ll
MOB00494
- bar (neg)
+ bar (pos)
MOB00468
- bar (neg)
+ bar (pos)