MOBILETT Plus / - E/ - M SPR8-215.898.01 Page 4 of 6 Siemens-Elema AB
Rev. 05 01.02 SPS-UD Solna, Sweden
7 - 4 Capacitor bank
• Remove the main bars (Fig. 5 for MOBILETT Plus/ Plus M and Fig. 6 for MOBILETT
Plus E).
• Remove the capacitor module.
Fig. 5 MOBILETT Plus/ Plus M Fig. 6 MOBILETT Plus E
Fig. 7
Residual voltage up to 353 VDC may still be present.
Risk of high voltage.
Use insulated tools and do not touch bars that may contain volt-
age (Fig. 7).
MOB00496
+ bar (pos)
- bar (neg)
MOB00495