MOBILETT Plus/ - E/ - M SPR8-215.898.01 Page 4 of 6 Siemens-Elema AB
Rev. 05 01.02 SPS-UD Solna, Sweden
8 - 4 Miscellaneous
• Remove the cable winder assembly.
• Remove the two cable ties (pos. 1 / Fig. 4).
• Remove four screws and four nuts (pos. 2 / Fig. 4) and Y-plate (pos. 3 / Fig. 4).
• Disconnect cable from upper plate (pos. 1 / Fig. 5).
• Remove the distance ring (pos. 2 / Fig. 5) from the axle.
• Loose the brake carefully to loose the tension of the cable winder spring.
• Remove securing screw (pos. 1 / Fig. 7).
• Pull out the cable winder.
Fig. 4 Fig. 5
MOB00620
2
1
3
4
MOB00621
1
2
3
3