Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Error messages of the imaging system 25
Page 25 of 68
Medical Solutions
-125 OK,
0xeffb
#1:Error
#2:Sensor unit:Grid error
#3:Because of the grid error, this image may have a quality
problem.
[Cause]
Although a grid warning had been detected in the sensor A/D
board, imaging was continued. Thus, the image may have been
degraded.
[Operation after the reply key is pressed]
When the OK key is pressed, the screen returns to the normal
display without any further action.
0xeffb
"Grid was changed during exposure (n)(effb)"
n: Sensor index (1 to 4)
-126 OK #1:Error
#2:Sensor unit:Temperature error
#3:The internal temperature of the sensor unit exceeded the
temperature limit. Cannot capture images until the tempera-
ture goes below the limit.#4:TEMP OVER (n)(effa) n: Sensor
index (1 to 4)
[Cause]
The sensor temperature has exceeded 49°C. Thus, the error for
aborting the operation has been detected. The system cannot
be used until the temperature decreases, which will take several
tens of minutes.
[Operation after the reply key is pressed]
The sensors automatically enter the sleep mode. The host com-
puter does nothing. If you try to make the system ready, the
same error will occur again. Press [OK] to close the window, a
few minutes later operate again. Note: The error might be able
to be avoided by reducing the ambient temperature.
Code Reply
keys
Message