30 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 30 of 68
Medical Solutions
-403 OK #1:Error
#2:Incomplete end error
#3:Do you want to store captured images in the internal storage
in order to recover them later? (Name of the patient) (Number of
images)
[Cause]
The last image(s) was/were not queued due to an incomplete
previous imaging operation.
[Operation after the reply key is pressed]
When OK is selected, the recovery process will start. When
Cancel is selected, the last images will be deleted. In this case,
normal exposure for imaging will start without recovering the last
images. When CANCEL is selected, a confirmation message
will appear before the image is deleted.
-509 OK #1:Error
#2:External storage error:
#3:Selected directory or disk for storage can only be read.
[Cause]
The external storage medium for the stored data is read-only.
[Operation after the reply key is pressed]
The system returns to the state before the occurrence of the
error.
-512 OK #1:Error
#2:External storage error:
#3:Cannot recognize the removable disk. Set the disk properly
and retry.
[Cause]
No medium is found on the removable disk.
[Operation after the reply key is pressed]
The system returns to the state before the occurrence of the
error.
Code Reply
keys
Message