EasyManua.ls Logo

Siemens MOBILETT XP Digital - Technical Safety Checks (TSC)

Siemens MOBILETT XP Digital
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Error messages of the imaging system 31
Page 31 of 68
Medical Solutions
-513 OK #1:Error
#2:Media control error
#3:Operation has been canceled because the external storage
or directory is busy.
[Cause]
The external medium is busy deleting or writing data. This error
occurs for example when you try to eject the disk while the
external medium is busy.
[Operation after the reply key is pressed]
After confirmation, press OK.
-514 OK #1:Error
#2:External storage error:
#3:The exposed images cannot be stored because the size of
the folder partition is smaller than the exposed images. Use a
new disk, erase the folder contents or enlarge the folder size
and retry.
[Cause]
An image cannot be stored in a folder on the external storage
medium when the size of the folder is smaller than the image.
Enlarge the folder size in the system setting.
[Operation after the reply key is pressed]
Selecting RETRY, stores the image again, while selecting CAN-
CEL cancels storage of the image to the external storage
medium. When CANCEL is selected, a confirmation message
appears before the image is deleted.
Code Reply
keys
Message

Other manuals for Siemens MOBILETT XP Digital

Related product manuals