32 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 32 of 68
Medical Solutions
-517 OK #1:Error
#2:Internal storage error:
#3:Processed image cannot be stored, because the internal
storage memory capacity is insufficient.
[Cause]
The remaining space on the internal disk is insufficient.
[Operation after the reply key is pressed]
The system returns to the state before the occurrence of the
error.
-518 OK #1:Error
#2:Internal storage error:
#3:DICOM image cannot be stored, because the internal stor-
age memory capacity is insufficient.
[Cause]
There is not enough space for storing the DICOM image.
[Operation after the reply key is pressed]
The system returns to the state before the occurrence of the
error.
-519 OK #1:Error
#2:External storage error:
#3: The external storage memory capacity is insufficient. Retry
after making space on the disk.
[Cause]
The space on the external storage medium is insufficient.
[Operation after the reply key is pressed]
The system returns to the state before the occurrence of the
error.
Code Reply
keys