Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Error messages of the imaging system 33
Page 33 of 68
Medical Solutions
-520 OK #1:Error
#2:Internal storage error:
#3:Exposed image cannot be stored, because the internal stor-
age memory capacity is insufficient.
[Cause]
The remaining space on the internal disk is insufficient.
[Operation after the reply key is pressed]
The system returns to the state before the occurrence of the
error.
-521 OK #1:Error
#2:Queue file open error:
#3:Queue file is not found.
[Cause]
Failed to open the queue file. This error occurs when no queue
file exists.
[Operation after the reply key is pressed]
*With the current system, this error is unlikely to occur.
-599 OK #1:Error
#2:File open error:
#3:File is not found.
[Cause]
This error occurs when the system has failed to open the file
during RESEND.
[Operation after the reply key is pressed]
Simply press OK.The system returns to the state before the
occurrence of the error.