34 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 34 of 68
Medical Solutions
-601 OK #1:Error
#2:Resend error
#3:File may have been automatically deleted.
[Cause]
The image cannot be resent because it was deleted for some
reason during retrieval.
[Operation after the reply key is pressed]
The image being displayed on the re-print screen is made
invalid and the image can no longer be resent.
-5100 OK #1:Error
#2:Sensor unit:Detect error
#3:Check the sensor power unit or cable connection. To recover
the system, shut it down.
[Cause]
The power of the imaging unit is OFF, or the cable is discon-
nected or damaged.
[Operation after the reply key is pressed]
The system only requires confirmation. Press OK. Only the
problematic sensor is disabled.
-5001 OK #1:Error
#2:Sensor unit:Detect error
#3:Check the sensor unit cable connection. To recover the sys-
tem, shut it down.
[Cause]
The cable is not connected although it is set for the sensor.
[Operation after the reply key is pressed]
The system only requires confirmation. Press OK. Only the
problematic sensor is disabled.
Code Reply
keys
Message