Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Error messages of the imaging system 35
Page 35 of 68
Medical Solutions
Tab. 6 Customer Error Table
-5002 OK #1:Error
#2:Grid error
#3:Remove grid for this exposure.
[Cause]
This error occurs when grids are detected in the buck/cassette
setting where imaging is expected to be performed without
grids.
[Operation after the reply key is pressed]
Press OK after the check.
Error Code Error Message Remedy
-2 Insufficient tmp disk space. Out-
put the saved data or erase the
backed-up data.
Touch OK and delete images.
Touch OK and shut down the sys-
tem.
If these errors occur, contact a Sie-
mens representative or distributor.
-3 Insufficient tmp disk space. Out-
put the saved data or erase the
backed-up data.
-120 Detector Unit: Self Check Error:
Detected an Error.
-121 Detector Unit: Sequence Error.
Call a service person.
-122 Detector Unit: Control Error:
Detected Control Error. Call a
service person.
-123 Detector Unit: PLD Error:
Detected PLD Error. Call a ser-
vice person.
-125 Detector Unit: Grid Error: Indi-
cates that this image may have a
quality problem.
Touch OK and shut down the sys-
tem.
-126 Detector Unit: Temperature
Error: The internal temperature
of the detector unit has
exceeded the temperature limit.
Touch OK and try again after a
while.