EasyManua.ls Logo

Siemens MOBILETT XP Digital - Page 44

Siemens MOBILETT XP Digital
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 44 of 68
Medical Solutions
-510 RETRY
CANCEL
#1:Warning
#2:External storage error:
#3:Selected directory or disk for storage does not exist. Or an
image cannot be saved on external storage, because the exter-
nal storage or directory is not properly set.
[Cause]
There is no directory or disk in the destination. Or recording can-
not be performed due to system function error.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored again. When
CANCEL is selected, the storing of the image onto an external
storage medium will be cancelled after a confirmation message
is displayed. Re-transfer is possible by sending the image from
the study list again.
-512 RETRY
CANCEL
#1:Warning
#2:External storage error:
#3: Cannot recognize the removable disk. Set the disk properly
and retry.
[Cause]
Disk is not inserted.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored. When CAN-
CEL is selected, the storing of the image will be cancelled after a
confirmation message is displayed. Re-transfer is possible by
sending the image from the study list.
Code Reply
keys
Message