Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Error messages of the imaging system 45
Page 45 of 68
Medical Solutions
-514 RETRY
CANCEL
#1:Warning
#2:External storage error:
#3:The exposed images cannot be stored because the size of
the folder partition is smaller than the exposed images. Use a
new disk, erase the folder contents or enlarge the folder size
and retry. The size of the folder partition: xxxxMB) (The size of
the exposed images: xxxxMB)
[Cause]
Images cannot be stored in the folder when the folder size of the
external storage medium is smaller than a study.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored. When CAN-
CEL is selected, the storing of the image will be cancelled after a
confirmation message is displayed. Re-transfer is possible by
transferring the image from the study list.
-515 RETRY
CANCEL
#1:Warning
#2:Internal storage error:
#3:Selected directory for storage is read-only.
[Cause]
The external storage medium is read-only.Use a writable exter-
nal storage medium.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored on the inter-
nal disk.When CANCEL is selected, the storing of the image will
be cancelled after a confirmation message is displayed.
Re-transfer is possible by sending the image from the study list.
Code Reply
keys
Message