46 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 46 of 68
Medical Solutions
-516 RETRY
CANCEL
#1:Warning
#2:Internal storage error:
#3:Selected directory for storage does not exist, or the path is
incorrect.
[Cause]
There is no directory in the destination, or recording cannot be
performed due to a system function error.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored on the inter-
nal disk.When CANCEL is selected, the storing of the image will
be cancelled after a confirmation message is displayed.
Re-transfer is possible by sending the image from the study list.
-517 RETRY
CANCEL
#1:Warning
#2:Internal storage error:
#3:Processed image cannot be stored, because the internal
storage memory capacity is insufficient.
[Cause]
The space on the internal disk is insufficient.
[Operation after the reply key is pressed]
The QA process will be performed again.
-518 RETRY
CANCEL
#1:Warning
#2:Internal storage error:
#3:DICOM image cannot be stored, because the internal stor-
age memory capacity is insufficient.
[Cause] There is not enough space to store the DICOM image.
[Operation after the reply key is pressed] When RETRY is
selected, image will be stored on the internal disk.When CAN-
CEL is selected, the storing of the image will be cancelled after a
confirmation message is displayed.