Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Error messages of the imaging system 47
Page 47 of 68
Medical Solutions
-519 RETRY
CANCEL
#1:Warning
#2:External storage error:
#3: The external storage memory capacity is insufficient. Retry
after making space on the disk.
[Cause]
There is not enough space on the external storage medium.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored on the exter-
nal storage medium.When CANCEL is selected, the storing of
the image will be cancelled after a confirmation message is dis-
played.
-520 RETRY
CANCEL
#1:Warning
#2:Internal storage error:
#3:Exposed image cannot be stored, because the internal stor-
age memory capacity is insufficient.
[Cause]
Space on the internal disk is insufficient.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored on the inter-
nal disk.When CANCEL is selected, the storing of the image will
be cancelled after a confirmation message is displayed.
-6000 OK #1:Warning
#2:Calibration error
#3:STAND(SENSOR-ID:XXX) Time has elapsed since the last
calibration. Calibrate now.
[Cause]
This error occurs when calibration has not been performed
within the set time.
[Operation after the reply key is pressed]
The calibration screen appears to prompt you to perform calibra-
tion.
Code Reply
keys
Message