Siemens AG SPR8-230.840.30.03.02 MOBILETT XP Digital
12.05 CS SD 24
Error messages of the imaging system 49
Page 49 of 68
Medical Solutions
-507 DICOM Printer Status Error: FD
has aborted DICOM transmis-
sion because of a FAILURE in
DICOM printer status. Check
receiving device.
• After eliminating the printer error,
touch RETRY.
• Or touch Cancel and cancel the
transfer.
-509 External Storage Error: Selected
directory or disk for storage is
read-only.
• Replace the disk with a writable disk
and select RETRY.
• Or select Cancel. Or contact the
network administrator.
-510 External Storage Error: Selected
directory or disk for storage does
not exist. Or image cannot be
saved on external storage,
because the external storage or
directory is not properly set.
-512 External Storage Error: Cannot
recognize the external storage.
Check disk settings and retry.
-514 External Storage Error: Exposed
images cannot be stored
because capacity of folder parti-
tion is smaller than size of
exposed images. Use a new
disk, erase folder contents, or
increase folder size and retry.
-515 Internal Storage Error: Selected
directory for storage is
read-only.
• Touch OK and shut down the sys-
tem.
• If these errors occur, contact a Sie-
mens representative or distributor.
-516 Internal Storage Error: Selected
directory for storage does not
exist, or attribution is incorrect.
-517 Internal Storage Error: Pro-
cessed image cannot be stored
because capacity of internal
storage memory is insufficient.
• Touch RETRY after the green mark
has disappeared from the message
area of the exposure screen.
-518 Internal Storage Error: DICOM
image cannot be stored because
capacity of internal storage
memory is insufficient.
• Select RETRY to save the image to
the disk again. Or select Cancel.
-519 External Storage Error: Capac-
ity of external storage memory is
insufficient. Retry after com-
pressing data.
• Replace the disk with a writable disk
and select RETRY.
• Or select Cancel. Or contact the
network administrator.
Error Code Error Message Remedy