50 Error messages of the imaging system
MOBILETT XP Digital SPR8-230.840.30.03.02 Siemens AG
12.05 CS SD 24
Page 50 of 68
Medical Solutions
-520 Internal Storage Error: Exposed
image cannot be stored because
capacity of internal storage
memory is insufficient.
• Select RETRY to save the image to
the disk again.
• Or select Cancel.
-6000 Calibration Error: *** have
passed since the last calibration.
Perform calibration.
• Touch OK and calibrate the unit.
Error Code Error Message Remedy