Technical specifications
11.4 Technical specifications of analyzer modules
Quick Start
Compact Operating Instructions, 05/2018, A5E31805656-06
123
11.4
Technical specifications of analyzer modules
11.4.1
OXYMAT 7 technical specifications
Table 11- 12 OXYMAT 7: General technical specifications
General
OXYMAT 7
OXYMAT 7 high-temperature
Information on the smallest
span
See module nameplate or device certificate
Max. power
consumpti-
on
Warm-up pha-
se
60 W 120 W
Weight (only module)
< 5.5 kg
Electrical safety In accordance with
DIN EN 61010-1
Table 11- 13 OXYMAT 7: Measuring ranges
Measuring ranges
OXYMAT 7
OXYMAT 7 high-temperature
Number of measuring ranges Max. 4; parameters can be
assigned freely
Max. 4; parameters can be as-
signed freely
Parameters can
be assigned in
the measuring
ranges
Smallest
possible
spans
0.5%, 1%, 2% or 5% O
2
0.5% O
2
for T < 72 °C
2% O
2
for 72 °C < T < 130 °C
Largest pos-
sible span
100% O
2
100% O
2
Note
Temperature change for high-temperature OXYMAT 7
If you change the temperature range, the device has to be re
-
calibrated. Detailed information
can be found in the operating manuals, see "
Bibliography" (Page 135).

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIPROCESS GA700 and is the answer not in the manual?

Siemens SIPROCESS GA700 Specifications

General IconGeneral
BrandSiemens
ModelSIPROCESS GA700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals