Connecting
5.2 Remote installation
SITRANS F M MAG 3100
36 Operating Instructions, 06/2010, A5E03005599-01
2. Remove SENSORPROM
®
unit from sensor and mount it on connection plate in
transmitter, see relevant transmitter operating instructions.
3. Fit the ½" NPT or M20 cable glands for supply and output cables.
4. Fit and connect electrode and coil cables as shown below.
6KLHOG
CAUTION
Unscreened cable ends
Keep un
screened cable ends as short as possible.
CAUTION
Prevent interference
Separate electrode and coil cables to prevent interference.
WARNING
Use in hazardous area
Connect mains protective earth to the PE terminal in accordance with the above
diagram in order to obtain potential equalization.