Troubleshooting/FAQs
7.2 Fluctuating process values
SITRANS F M MAG 3100
Operating Instructions, 06/2010, A5E03005599-01
45
7.2 Fluctuating process values
Question
Why do the displayed process values fluctuate when the electrode cable is moved?
Answer
There are several causes to the fluctuating process values:
Deposits on electrodes
Clean the electrodes.
Defect electrode cable
Replace the cable
Incorrect cable connection
Connect the electrode cable (82, 83, 0 and shield) according to the instructions in
chapter Connecting (Page 33)
NOTICE
Vibrating environments
It is recommended to use special low noise cables for sensor sizes DN 2 and 3
installed in vibrating environments.