Safety notes
2.3 Installation in hazardous area
MAG 6000 I / MAG 6000 I Ex de transmitter
14 Operating Instructions, 04/2016, A5E02083319-AG
Compact transmitter version:
● Ex d e [ia] ia IIC T3-T6 Gb
Ex tD A21 IP67 T** °C (** pipe temperature +5K, but not less than T85 °C
(MAG 3100 / MAG 3100 P)
● Ex d e [ia] ia IIB T3-T6 Gb
Ex tD A21 IP67 T** °C (** pipe temperature +18K, but not less than T85 °C
(MAG 1100 / MAG 1100 F)
For sensors:
Refer to sensor certificates:
MAG 3100 / MAG 3100 P: Sira 07ATEX1182X (DN 15-DN300); Sira 03ATEX3339X (DN350-
DN2000)
MAG 1100 / MAG 1100 F: Sira 07ATEX3181X
Safety parameters associated with the remote transmitter version
Table 2- 1 Intrinsically safe data
Sensor electrode input (Terminals 0, 81, 82, 83, 84 - "ia circuits")
O
O
O
115 to 230 VAC or 18 to 30 VDC
Max. allowable supply voltage U
m
(according to
264 V
Ambient temperature -20 to +60 °C (-4 to 140 °F)