Technical data
11.4 Output characteristics
MAG 6000 I / MAG 6000 I Ex de transmitter
Operating Instructions, 04/2016, A5E02083319-AG
65
Current output
As pulse output ± (0.1% of actual flow + 0.05% FSO)
Effect of ambient temperature
Display/frequency/pulse output
Current output
< ± 0.003% / °C act.
< ± 0.005% / °C act.
Effect of supply voltage
< 0.005% of measuring value on 1% change
Repeatability ± 0.1% of actual flow for V ≥ 0.5 m/s (1.5 ft/s) and
conductivity ≥ 10 μS/cm
Additions in the event of deviations from reference conditions
11.4
Output characteristics
Current output
Output characteristics
Bidirectional mode
Unidirectional mode
4 to 20 mA
Frequency output
Bidirectional mode
Unidirectional mode
Pulse output
Bidirectional mode
Unidirectional mode

Table of Contents