Commissioning
6.2 Menu structure
MAG 6000 I / MAG 6000 I Ex de transmitter
Operating Instructions, 04/2016, A5E02083319-AG
41
Keypad
The keypad is used to set the flowmeter. The keys function as follows:
TOP UP KEY
This key (when held for 2 sec.) is used to switch between
operator menu and setup menu. In transmitter setup menu, a
short press will cause a return to previous level.
FORWARD KEY
This key is used to step forward through the menus. It is the
only key normally used by the operator.
BACKWARD KEY
This key is used to step backwards
through the menus.
CHANGE KEY
With this key settings or numerical values are changed.
SELECT KEY
With this key figures to be changed are selected.
LOCK/UNLOCK
KEY
This key enables the operator to change settings and it gives
access to submenus.
6.2
Menu structure
The menu is built up of two parts. An
operator menu
and a
setup menu
, see also Menu
diagrams (Page 73).
Operator menu
The operator menu is for daily operation. It can be customized in the operator menu setup in
which the menus that are to be available to the operator are selected. The transmitter always
starts up in operator menu No. 1. The forward
and the backward keys are used to
step through the operator menus.
Setup menu
The setup menu is for commissioning and service only. Access to the setup menu is gained
by pressing the top up key
for 2 seconds. The setup menu operates in two modes:
View mode
Setup mode
View mode
is a read-only mode. The pre-selected settings can only be scanned.
Setup mode
is a read and write mode. The pre-selected settings can be scanned and
changed. Access to the setup mode is password-protected. The factory set password is
1000.
Access to a submenu in the setup menu is gained by pressing the lock key . Press the
top up key
briefly to return to the previous menu. Press longer (2 sec.) to exit the setup
menu and return to operator menu no. 1.