11.6 Cable requirements
MAG 6000 I / MAG 6000 I Ex de transmitter
68 Operating Instructions, 04/2016, A5E02083319-AG
Technical data
Standard cable
(electrode/coil)
Double-shielded cable
(electrode)
Basic data
No. of conductors
3
3
Sqr. area
1.5 mm
2
0.25 mm
2
Screen
Yes
Double
Color code
Brown, blue, black
Brown, blue, black
Outside color
Grey
Grey
Ext. diameter
7.8 mm
8.1 mm
Conductor
Flexible CU
Flexible CU
Isolation material
PVC
PVC
Ambient temperature Flexible installation -5 ... +70°C
(23 ... 158°F)
-5 ... +70°C
(23 ... 158°F)
Non-flexible installation -30 ... +70°C
(-22 ... 158°F)
-30 ... +70°C
(-22 ... 158°F)
Cable parameter
Capacity
161.50 pF/m
-
Inductance
0.583 μH/m
-
L/R
43.83 µH/Ω
-
For more information on cable lengths, empty pipe detection, and conductivity, see
Operating Instructions for relevant sensor.
11.6
Cable requirements
Coil
cable
Electrode
cable
Basic data
No. of conductors
2
3
Min. sqr. area
0.5 mm
2
0.2 mm
2
Screen
Yes
Yes
Max. capacitance
N/A
350 pF/m
Max. cable loop resistance
Media temperature:
< 100 °C
40 Ω
N/A
> 200 °C
6 Ω
N/A
Cable glands on sensor
M20x1.5 gland - Cable ø 5 ... 13 mm (0.20 ... 0.51 inches)
½ NPT gland - cable ø 5 ... 9 mm (0.20 ... 0.35 inches)