Roughness, menu 7.1.3.2 (to be changed for SONOKIT only)
Roughness is the value for the surface inside the pipe.
The range of this value is defined to 0.01 mm to 10.0 mm because standard Siemens sensors
have a roughness of approximately 0.4 mm.
Smooth plastic pipe 0.1 mm
Polished stainless steel pipe 0.1 mm
Standard carbon steel pipe 0.4 mm
Rusty carbon steel pipe 1...2 mm
Concrete pipe 2...5 mm
Tracks, menu 7.1.4 (to be changed for SONOKIT only)
This menu is for the parameters which are for the track settings. In the following table only the
parameters for track 1 are explained.
The parameters for tracks 2, 3, and 4 are to be handled like track 1.
Note
The menus 7.1.4.1.x.x and 7.2 are not for flowmeter systems with sensors
SONO 3100 and SONO 3300. Please do not change these values. They are determined
and preset at the factory.
Track 1, menu 7.1.4.1.1 and menu 7.1.4.2.1 (to be changed for SONOKIT only)
For SONOKIT:
Data to be entered from “SENSOR GEOMETRY MEASUREMENT REPORT”
Length 1
7.1.4.1.1 Length 1 The length of the first track 0.2 m >0 m ... 8.0 m
This parameter is the distance between the ultrasonic transducers, as illustrated in the picture.
The value is necessary in order to calculate the ultrasonic velocity from the time of flight and
the distance between the transducers.
Angle 1, menu 7.1.4.1.2 and menu 7.1.4.2.2 (to be changed for SONOKIT only)
Angle of track 1.
It is not necessary to enter this menu is not necessary for flowmeter systems with sensors SONO
3100 and SONO 3300. These values are determined and preset at the factory.
For SONOKIT: Data to be entered from “SENSOR GEOMETRY MEASUREMENT REPORT”