8 TECHNICAL DATA
104
SITRANS FX330
www.siemens.com/flow 08/2022 - A5E36751269-AG EN
8.1 Functional principle
Vortex flowmeters are used to measure the flow of gases, vapours and liquids at completely
filled pipes.
The measuring principle is based on the Karman vortex street. The measuring tube contains a
bl
uff body at which vortex shedding occurs and which is detected by a sensor unit located behind.
The frequency f of the vortex shedding is proportional to the flow velocity v. The non-dimensional
St
ouhal number S describes the relationship between vortex frequency f, width b of the bluff
bo
dy and the average flow velocity v:
The vortex frequency is recorded at the flow sensor and evaluated at the signal converter.
Figure 8-1: Functional principle
=
f
S
.
v
b