3 INSTALLATION
30
SITRANS FX330
www.siemens.com/flow 08/2022 - A5E36751269-AG EN
3.4.2 Installation when measuring steam and gases
Figure 3-4: Recommended installation
1 Installing the device in an upper pipe bend
2
If the device is installed in a downpipe, a downpipe must be installed immediately after it
Figure 3-5: No
t recommended installation
1 Lower pipe bends
2
Condensate
CAUTION!
Installing the device in a lower pipe bend: there is a risk of condensate forming.
Condensate can lead to cavitation and inaccurate measurement. Under certain circumstances
th
e device can be destroyed and the measured medium can leak.