A5E33701270 COMMUNICATIONS FOR SITRANS LUT400 (HART) – MANUAL Page 31
mmmmm
Operation - PDM
6) Once Auto TVT is complete, click on Transfer to Device. To exit, click on Close. Auto
TVT is enabled and the learned TVT will be used.
7) To turn Auto False Echo Suppression off or on, reopen the Auto False Echo
Suppression window, change the Auto False Echo Suppression to On or Off, and
click on Transfer to Device.
Before Auto False Echo Suppression
default TVT
TVT
Hover
Level
Level (db)
echo marker
After Auto False Echo Suppression
material level
false echo
Auto False Echo Suppression
Range
echo
marker
Level (db)
default
TVT
material level
false echo
learned TVT