Settings
39 / 60
Parameter Values Default value
«LED-Temp» On/ Off Off
Storage of LED temperature.
«Humidity» On / Off Off
Storage of humidity value.
9.4.12 Menu: System
Parameter Values Default value
«Mandatory operation» 60 s ... 60000 s 900 s
Time period after which the device automatically returns to measuring mode without manipulation (mandatory
operation). This prevents the measuring device from remaining in service mode for any length of time and no
relevant measured value/ limit value being output.
«Date format» DD.MM.YYYY/
DD/MM/YYYY/
MM/DD/YYYY
DD.MM.YYYY
Set the format of the date.
«Summer time» No/ Yes/ Europe Europe
Set daylight saving time. For Europe, daylight saving time is set on the last Sunday in March and winter time on
the last Sunday in October.
«OTA update transmits extended
diagnostic data»
On/ Off On
During an online firmware update, operating hours, temperatures, voltages, intensities of the light sources and
the error history are transmitted.
«Contact information» ... Sigrist-Photometer AG
Enter line 1 of the contact information (max. 47 characters).
«Contact information» ... Switzerland
Enter line 2 of the contact information (max. 47 characters).
«Contact information» ... +41 41 624 54 54
Enter line 3 of the contact information (max. 47 characters).
«Contact information» ... info@sigrist.com
Enter line 4 of the contact information (max. 47 characters).
9.4.13 Menu: Meas. Channels
Meas. Channels\ Channel C1 ... Cn
Parameter Values Default value
«Linearisation» [Define...] 1100/1100 - 0.000/0.000
Definition of a customer-specific linearisation with eight interpolation points (actual/nominal value pairs). Mea-
sured values are linearly interpolated between basic values. Measured values that are smaller than the smallest
nominal basic value are treated as the smallest basic value. Measured values outside the highest nominal basic
value are displayed as an overflow (****).