Settings
31 / 60
Parameter Values Default value
«Outputs» Off/ All Off/ All On/ 1 ... n On Off
Output specific states at digital outputs.
«Light source» Off/ 1, 2 Off
Manually switch the light source on or off for test purposes or fault location (1 = LED turbidity, 2 = LED colour
(optional)).
9.3.3 Menu: Recalibration
Submenu: Recalibration C1 ... Cn
Parameter Values Default value
«Nominal value» Device-specific -
Enter the value of the control unit belonging to the unit or the nominal value for the formazine solution.
«Actual value» Current measured value -
Current measured value.
«Adjustment» Trigger... -
Triggers the adjustment. Calculates a new correction factor from the actual and nominal values.
«Act.corr» 0.500 ... 2.000 1.000
Specifies the current correction factor, which corrects the deviation from the factory calibration.
9.3.4 Menu: History
History\ Fault
Parameter Values Default value
«Fault» - -
View recorded warnings, faults, prio faults and information.
History\ Adjustment
Parameter Values Default value
«Adjustment» - -
View chronologically recorded adjustment values.
9.3.5 Menu: System info
Parameter Values Default value
«Device type» Device name
View the device type.
«Serial number» Device-specific
View the serial number. This number is important when contacting customer service.