Servicing
46 / 60
10.3 Clean sensor head
WARNING!
Do not remove the photometer without first draining
the process line!
u
Remove the photometer.
u
Clean windows (1), (2), (3) with mild, abrasive-free
cleaning agent (e.g. alcohol or soap). Use a soft,
non-lint cloth.
u
Install the photometer.
1
3
2
10.4 Calibration check
Calibration check, general
l
An adjustment leads to deviations from the previous measured value.
l
The turbidity measuring channels (C1, C2) can be calibrated with the solid reference and tap water or for-
mazine.
l
For the contamination and colour channel (C3, C4), distilled water is used.
DANGER
Skin or eye damage caused by formazine (hydrazine sulphate).
Unprotected skin or eye contact with formazine may cause skin or eye damage.
u
Wear protective goggles and gloves.
u
Wash hands after work.
10.4.1 Overview of control unit
(1) TurBiScat PM 40
(2) Clamp ring
(3) Adapter VARINLINE
®
(4) Level indicator
(5) Filling funnel holder
(6) Filling funnel
1
2
3
4
5 6
Overview of control unit

Table of Contents

Other manuals for SIGRIST TurBiScat PM 40

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SIGRIST TurBiScat PM 40 and is the answer not in the manual?

SIGRIST TurBiScat PM 40 Specifications

General IconGeneral
BrandSIGRIST
ModelTurBiScat PM 40
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals