Troubleshooting
52 / 60
Code Message Description Possible causes
W27 ADJUSTMENT Adjustment of the device could not
be carried out
l
Device is soiled
l
Nominal value for adjustment
does not match the value of the
medium
W28 SENSOR CHECK Automatic sensor check failed
l
Too much external light near the
measuring cell (e.g. sight glass)
l
Device open
l
Defective optics/electronics
Contact service engineer
W29 OVERTEMP Temperature in the device has ex-
ceeded 65 °C
l
Medium or ambient temperature
too high
l
No or defective cooling unit
W30 HUMIDITY Relative humidity in the device
above the set limit value
l
Desiccant is saturated
l
Housing seal defective
l
Device was open too long
W33
...
W40
CURRENT 1...8 Current output is disturbed
l
Open connection terminals
l
Interruption in the current loop of
the measured value output
Contact service engineer
W41 TEMP.SENSOR Inner temperature sensor has failed
l
Defective electronics
Contact service engineer
W43 EXTERNAL ON An external event is signalled via a
digital input
l
External fault
W53 IO_PORT Communication interruption to touch
screen
l
Defective electronics
Contact service engineer
W57 ABSORPTION TOO
HIGH
Light beam blocked by the measur-
ing cell
l
Very dark medium or foam
W78 SERVICE Indicates when maintenance is due
l
Contact service engineer
11.4 Fault messages
The following fault messages may be displayed during operation.
Code Message Description Possible causes
E8 SERIAL 1 Communication interruption be-
tween main controller and sensor
board
l
Defective electronics
Contact service engineer
E16 U ANALOG One of the internal analogue volt-
ages is outside the permissible
range
l
Defective electronics
Contact service engineer
E17 MEASURING FAULT Measurement value acquisition is
disturbed
l
Instrument not in process line
l
Air bubbles present in the medium
l
Extraneous light near the measur-
ing point (e.g. sight glass)
l
Defective electronics
Contact service engineer
E19 LIGHT SOURCE 1 Detector for monitoring the light
source is not receiving light from the
corresponding light source.
l
Defective light source
Contact service engineer

Table of Contents

Other manuals for SIGRIST TurBiScat PM 40

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SIGRIST TurBiScat PM 40 and is the answer not in the manual?

SIGRIST TurBiScat PM 40 Specifications

General IconGeneral
BrandSIGRIST
ModelTurBiScat PM 40
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals