INSTRUCTION MANUAL
LabScat 2
Laboratory Turbidimeter

Table of Contents

Questions and Answers

  • R
    russellmorenoSep 9, 2025
    What does LED1 FAILURE mean in SIGRIST LabScat 2?
    • H
      Heather EdwardsSep 9, 2025
      LED1 FAILURE in SIGRIST Measuring Instruments indicates a defective light source.
  • A
    Alexandria JacksonAug 23, 2025
    How to troubleshoot no reading on SIGRIST LabScat 2 Measuring Instruments?
    • D
      David WellsAug 23, 2025
      If there is no reading on your SIGRIST Measuring Instruments, check whether the mains device is plugged in and whether the supply voltage is present. Also, check whether the instrument is switched on.
  • C
    christopher29Aug 25, 2025
    What causes WATER SHORTAGE in SIGRIST Measuring Instruments?
    • S
      sjacksonAug 25, 2025
      A WATER SHORTAGE message on SIGRIST Measuring Instruments can be caused by several factors. The regulator valve might be closed, or the water connection might not be correct. This message can also briefly appear if a large bottle has been removed from the water bath. Another cause can be that the float switch is stuck, or there may be a defect in the electronic system.
  • B
    Bryan HughesAug 26, 2025
    What causes ADJUSTFAULT error in SIGRIST LabScat 2 Measuring Instruments?
    • K
      Kevin AnthonyAug 27, 2025
      An ADJUSTFAULT error with SIGRIST Measuring Instruments can occur if the instrument is soiled or if the set point for the adjustment does not match the value of the sample.
  • G
    Gabriel ConnerAug 28, 2025
    What causes MEASUR.FAULT in SIGRIST Measuring Instruments?
    • N
      Nancy RojasAug 28, 2025
      A MEASUR.FAULT error in SIGRIST Measuring Instruments can be caused by a very unstable sample, large air bubbles in the water bath, strong external light influence, an open water bath cover, or a defect in the electronic system.
  • S
    shelbylopezAug 29, 2025
    Why is there a HIGH STANDARD DEVIATION in SIGRIST LabScat 2 Measuring Instruments?
    • K
      Kyle GregoryAug 29, 2025
      A high standard deviation in SIGRIST Measuring Instruments can be caused by an unstable sample, which may stem from particles or gas emissions.
  • M
    mbassSep 18, 2025
    What causes CRC DISPLAY error in SIGRIST Measuring Instruments?
    • D
      Diana BlairSep 18, 2025
      A CRC DISPLAY error in SIGRIST Measuring Instruments can be caused by electromagnetic malfunctions or a defect in the electronic system.
  • M
    Mark BautistaSep 19, 2025
    What does EXT RAM error mean in SIGRIST Measuring Instruments?
    • J
      Juan BrownSep 19, 2025
      An EXT RAM error in SIGRIST Measuring Instruments indicates a defect in the electronic system.
  • T
    Tyler RodriguezSep 7, 2025
    What causes TOO MUCH EXTERNAL LIGHT error in SIGRIST Measuring Instruments?
    • R
      Rhonda MorrowSep 8, 2025
      TOO MUCH EXTERNAL LIGHT error in SIGRIST Measuring Instruments is caused by the water bath cover being open.
  • M
    Mr. Kenneth RoySep 4, 2025
    What does OVER TEMP mean on SIGRIST LabScat 2 Measuring Instruments?
    • J
      jolsonSep 4, 2025
      An OVER TEMP error in SIGRIST Measuring Instruments indicates that the medium or ambient temperature is too high, or that there is defective or no cooling.

Summary

General User Information

Safety Symbols and Pictograms

Explains danger symbols, safety warnings, and pictograms used in the manual.

Instrument Overview

General Safety Points

Operational Dangers and Residual Risks

Addresses hazards during instrument operation and any remaining risks after safety measures.

Assembly

Setup and Connections

Guides on selecting an operating location and connecting hoses for the water bath.

Electrical Installation

Power Connection Safety

Provides critical safety instructions for connecting the instrument to the power supply.

Commissioning

Operation

Basic Operation and Touchscreen Use

Introduces fundamental operation principles and how to interact with the touchscreen interface.

Measurement and Data Handling

Covers logging data, performing measurements, copying data, and switching to service mode.

Servicing

Troubleshooting

Malfunction Identification and Resolution

Offers guidance on identifying and troubleshooting common instrument malfunctions and errors.

Warning and Error Message Interpretation

Explains the meaning of various warning messages, error codes, and prioritized faults.

Customer Service Information

Decommissioning and Storage

Packaging, Transport, and Returning

SIGRIST LabScat 2 Specifications

General IconGeneral
Protection ClassIP65
Resolution0.001 NTU
Measuring PrincipleNephelometry
Measuring Range0 to 4000 NTU
Sample Temperature0 to 50 °C
Interfaces4-20 mA
Storage Temperature-20 to 60 °C
Accuracy±2% of reading or ±0.01 NTU, whichever is greater