10.4 Prioritized error messages and effect
When there is a prioritized fault, the cause of the malfunction is serious.
If a prioritized fault occurs during operation, it has the
following effects:
The measuring values go to 0.
Prioritized faults can be cleared only by a service
engineer.
When the Prio message occurs, the color of the
status display changes to red and the error text
describes which prioritized fault it is.
Example: PRIO DEFAULT VALUES
The following prio error messages can be displayed:
The default values were load-
ed.
If no parameters were initial-
ized or if all parameters were
lost, the default values are
loaded.
A fault was determined when
the expert data was checked.
Electromagnetic malfunctions.
Defect in the electronic system.
A fault was determined when
the user data was checked.
Electromagnetic malfunctions.
Defect in the electronic system.
A fault was determined when
the display data was checked.
Electromagnetic malfunctions.
Defect in the electronic system.
A fault was determined when
the RAM in the graphic con-
troller was checked.
Defect in the electronic system.
Table 5: Possible prio error messages