Instruction Manual LabScat 2 Servicing
40 12112E/2
9 Servicing
9.1 Servicing schedule
WHEN
WHO
WHAT
PURPOSE
As needed
Operator
Cleaning the water
bath.
Section 9.2
Obligatory measure for main-
taining measuring precision.
Interval dependent on water
quality and handling.
Every three
months or as
needed
Operator
Adjustment with
checking unit.
Section 9.3
Measure for maintaining meas-
uring precision.
For warning
humidity
Section 7.3
Operator
Changing the desic-
cant.
Section 9.4
Obligatory measure for mainte-
nance of the instrument.
Table 1: Servicing schedule
9.2 Cleaning the water bath
Soiling of the water bath is generally compensated by the LabScat 2. After a certain operat-
ing duration, however, depending on the operating conditions and the water quality, soiling
can become so extreme that it can no longer be compensated.
The following procedure describes water bath cleaning:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Switch off the instrument in accordance with
Section 5.2.
2.
Empty the water bath in accordance with Sec-
tion 4.4.
3.
Remove the water bath cover:
Pull the two side locks outward at the same
time and lift the water bath cover.