Instruction Manual LabScat 2 Assembly
20 12112E/2
4.3 Filling the water bath
The regulator valve must be pressed in completely to fill the water bath.
The water bath should not be without water circulation for a long period of time.
The continual water circulation ensures that the water bath is always filled, also for dif-
ferent bottle sizes.
The water bath must be cleaned periodically.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Fully open the regulator valve counter clock-
wise and fill until the water overflows.
2.
Set the regulator valve to approx. 0.5 l/min.
4.4 Emptying the water bath
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Stop the water supply by closing the regulator
valve.
2.
Pull out the regulator valve to the stop so that
the water can flow out of the water bath.