Operation Instruction Manual LabScat 2
12112E/2 27
7.5 Functions of the log screen (Log button)
The file can also be viewed via the web interface as a text file.
The log file is deleted after it has been transferred to a USB stick.
When exiting the log screen, the file is shortened to a maximum of 1000 entries.
Figure 8: Functions of the Log display

Date, time and storage location

Fault messages and warning messages

Description of the measurement

Lin curve used

Measuring values 90°/25° with minimum value, maximum value and standard devia-
tion.
</>: Forward/back one storage location
<</>>: Forward/back 20 storage locations
ESC: Exit logger
7.6 Perform measurement
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Fill the water bath in accordance with Section
4.3.
Depending on the water qua-
lity, air bubbles may appear in the
water bath. Stirring can improve
the situation.
2.
Put the sample in the water bath and close the
water bath cover.