Instruction Manual LabScat 2 Servicing
42 12112E/2
WORKSTEP
ADDITIONAL INFO / IMAGES
7.
Remove lime deposits from the inside of the
water bath with a commercially available de-
calcifier.
Rinse the water bath with clear water.
8.
Clean the window on the inside of the water
bath (arrow positions) with a clean, soft cloth
(e.g. a cotton cloth).
9.
Insert the floater into the guide bar (circle posi-
tion) and ensure that it can move freely.
10.
Mount and fasten the centering mechanism.
11.
Position the water bath cover on the locking
pins and press down until the locks snap into
place.