CM5_UM_ 9752090319_GB_V0_9 52 / 67
External remote mode is
enabled.
Send the generator for
repair, see section 10.
There is no bus
communication
A “ ” flashes on the
display
The bus master is not
operational.
Start up the bus master
or restart the system.
Incorrect bus address
set.
Set the correct bus
address.
Problems with cabling. Check the cabling,
connectors and the
terminating resistances
of the bus.
Generator is in
locked mode (“
”)
and will not unlock
The “ ” symbol
flashes on the screen
several times
The generator has been
locked by the
superuser.
Get the superuser to
unlock the generator
using the menu, see
menu 6110.
The display shows
the text "Settings
Eeprom ERROR!
Factory settings are
restored.” during
start-up
Error occurred whilst
saving information.
Depress the [ ] key
and restore the
generator settings. If
the problem persists,
send
the generator for
repair. See section 10.
Configuration
Eeprom ERROR!
appears on the
display during start-
up
An error has occurred in
the Eeprom.
Send the generator
away for repair (see
section 10).
Remote ADC input
error!” appears on
the display
The signal voltage is too
high on one of the
analogue control inputs.
Check whether the
analogue control signals
are correctly configured
(see menus 6630, 6640,
6650 and 6660).
Too many ARCs
error!” appears on
the display
Too many arcs have
occurred in too short a
time.
Reduce the set high
voltage.
Communication
error!” appears on
the display
A communication
problem has occurred
with the Anybus
module.