CGM User Guide
53 Section 3 Understand Your iCan i3 CGM Results
Rise or Fall Rate Alert
• Slowly Rise or Fall: Until conrmation vibrates and beeps 3 times
every 3 minutes with APP notication.
• Quickly Rise or Fall: Until conrmation vibrates and beeps 6 times
every 3 minutes with APP notication.
• Rapidly Rise or Fall: Until conrmation vibrates and beeps will be
continued.
Signal Loss Alert
This tells you when you are not getting iCan results. Your mobile device
may be too far from your transmitter or there may be something, such
as a wall or water, between your transmitter and your display device.
Unlike other alerts, Signal Loss cannot make a sound or vibrate, you
will receive APP notication for every 3 minutes until reconnected.
To x this problem, keep your transmitter and display device within 6
meters of each other. If that does not work, turn Bluetooth o and on.
Wait 10 minutes, if this still does not work, restart the mobile device and
reopen the CGM APP.
During signal loss, use your BG meter to check your glucose and make
any treatment decisions.