SITA Messtechnik GmbH - Operating Instructions SITA ConSpector
8
3 Quick-start
The quick-start chapter contains the most important information for operation, conduct-
ing a measurement and inspection of the device in abbreviated form. The single sections
of the quick-start chapter provide cross references to later chapters with more detailed
information on the respective topic.
3.1 Sampling
Take samples at a well-mixed and representative sampling location. Always carry out
measurements at the same sample temperature. Allow samples to cool down to room
temperature before starting the measurement.
3.2 Perform measurement
1. Switch on the SITA ConSpector (press -key for 2 s).
2. Choose the desired measuring profile. The active
measuring profile appears at the lower right edge of
the display. Use the arrow selection keys Liquid to
change the profile.
3. Place the stainless steel beaker filled with approx. 80 ml of the sample liquid (ap-
prox. 1 cm below the rim of the beaker) on the support surface of the magnetic
stirrer. Place a magnetic stick in the beaker and turn the rotary knob to mid posi-
tion.
4. To immerse the immersion probe into the sample liquid, lower the device holder
by using the star knob screw (see set-up in picture above). Immerse the immersion
probe approximately half way into the bath sample. The limiter should rest against
one side of the beaker.
5. Use the lid to cover the bath sample.
immersion
probe
limiter
immersion
depth
active profile
lid closed
12 o'clock position
measurement in
stainless beaker