SITA Messtechnik GmbH – Operating Instructions SITA ConSpector
A temperature sensor is integrated in the immersion probe to measure the temperature of the
sample.
4.3 Power supply
In mobile operation power is supplied by an integrated lithium ion battery. The integrated charging
circuit of the device charges the battery gently within a maximum of 7 hours. So express battery
servicing is not needed and you can operate the device with the power pack supplied or an USB
interface if possible. When the battery is empty (as shown on the battery charge display or device
cannot be switched on) connect the device with the power pack supplied or a USB port.
For initial use charge the battery completely using the power pack. If the battery capacity decreases
noticeably after about 1,000 charging cycles, it can be replaced (see section 8.4).
It is possible to disable the battery charge function (see section
4.6) and thus conserve the battery if you permanently use the
power pack or the PC interface to operate the SITA ConSpector.
If the charging function is disabled, a warning symbol will ap-
pear on the display (see section 4.4).
Stored measured values and set device parameters will be maintained even if the battery is empty
or when there is no power supply.
4.4 Display
The SITA ConSpector has a display on which the necessary information is output according to the
various operating states (FluoLevel, Cont(amination)Level or User-defined profile). The most im-
portant output contents are shown below as an introduction:
Toolbar:
Battery charge level (3 bars = fully charged)
PC or power supply connection
Signal tones On/Off
Warning "Charge battery off"
Operating mode
(here: FluoLevel)
Profile name and warnings