General
1-2
b) The output voltage of portable X-ray unit is 40-125kV
c) The output current of portable X-ray unit is 10-80mA
d) The small tube focus is 0.6mm.; The big tube focus is 1.8mm
1.5 Composition and Meaning of Model
SR - 8210
Product model
Product code, with 2 letters (from left to right) referring to:
S- Smart, R- Radiographic X-ray device
1.6 Environmental Conditions
1.6.1 Normal operation conditions
a) Ambient Temperature: 10℃~40℃.
b) Relative Humidity: 30%~75%.
c) Atmospheric Pressure: 700hPa ~1060hPa.
d) Ambient environment: no strong mechanical vibration and strong
electromagnetic interference, no corrosive gas or dust.
1.7 Power Requirement for Normal Operation
Battery Charger:
a) Rated voltage: 100-240V~;
b) Rated current: 1.2A (max);
c) Operating frequency: 50-60Hz;
d) Output voltage: 50.4V
;
e) Output current: 1.4A;
f) Class II equipment;
The power supply of the portable X-ray unit is internal power supply.